Contec Pulse Oximeter is used to measuring the pulse oxygen saturation and pulse rate through finger. This digital pulse oximeter is suitable for being used in family, hospital, oxygen bar, community healthcare, physical care in sports (It can be used before or after doing sports, and it is not recommended to use the device during the process of having sport) and etc.Principle of the CMS50D Pulse Oximeter is as follows: Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse Scanning & Recording Technology.
Contec Pulse Oximeter Features:
- Integrated with SpO2 probe and processing display module
- Small in volume, light in weight, and convenient in carrying
- Operation of the product is simple, low power consumption
- SpO2 value display
- Pulse rate value display, bar graph display
- Pulse waveform display
- The display mode can be changed
- Screen brightness can be changed
- Low-voltage indication: low-voltage indicator appears before working abnormally which is due to low-voltage
- Automatically power off function: when the device is under the state of measuring interface. it will automatically power off within 5 seconds if the finger falls out of the probe
- The display format can be saved after power off
- Standard Fingertip Pulse Oximeter
Fingertip Oximeter Specification
| Display Mode | 0.96″ Dual-color OLED display(blue and yellow) |
| SpO2 Measuring Range | 0%~100%, (the resolution is 1%). |
| Screen Resolution | 128*64 |
| PR Measuring Range | 30bpm~250bpm, (the resolution is 1bpm) |
| Accuracy | 70%~100%: ±2% ,Below 70% unspecified |
| Accuracy | ±2bpm or ±2% (select larger) |
| Measurement Performance in Weak Filling Condition: | SpO2 and pulse rate can be shown correctly when the pulse-filling ratio is 0.4%. SpO2 error is ±4%, pulse rate error is ±2 bpm or ±2% (select larger). |
| Resistance to surrounding light | The deviation between the value measured in the condition of man-made light or indoor natural light and that of the darkroom is less than ±1% |
| Power Consumption | less than 30mA |
| Voltage | DC 2.6V~3.6V |
| Power Supply | 1.5V (AAA size) alkaline batteries × 2 |
| Battery working hour | The minimum continually work time is 20 hours, the theoretical number is 36 hours |
| Safety Type | Interior Battery, BF Type |
RN: 002500S00695H02, CMS50D